8. Important information
Reorient or relocate the receiving antenna or equipment.
Increase the separation between the equipment and receiver.
Connect the equipment into an outlet on a circuit different from that to which the receiver is connected.
Consult the dealer or an experienced technician for help.
Electromagnetic immunity
The Coronis 3MP LED Display is intended for use in the electromagnetic environment specied below.
The customer or the user of the Coronis 3MP LED Display should assure that it is used in such an envi-
ronment.
Immunity test
IEC 60601
Test levels
Compliance level
Electromagnetic
environment
guidance
Electrostatic discharge
(ESD)
IEC 61000-4-2
± 6kV contact
±8kVair
± 6kV contact
±8kVair
Floors should be wood,
concrete or ceramic tile.
If oors are covered with
synthetic material, the
relative humidity should
be at least 30%
Electrical fast
transient/burst
IEC 61000-4-4
± 2kV for power supply
lines
± 1kV for input/ output
lines
± 2kV for power supply
lines
± 1kV for input/ output
lines
Mains power quality
should be that of a typical
commercial or hospital
environment
Surge
IEC61000-4-5
± 1 kV line(s) to line(s)
± 2 kV line(s) to earth
± 1 kV line(s) to line(s)
±2kVline(s)toearth
Mains power quality
should be that of a typical
commercial or hospital
environment
Voltage dips, short
interruptions and voltage
variations on power
supply input lines
IEC 61000-4-11
<5%U
T
1
(> 95% dip in
U
T
) for 0.5 cycle
40% U
T
(60%dipinU
T
)
for 5 cycles
70% U
T
(30%dipinU
T
)
for 25 cycles
<5%U
T
(>95% dip in
U
T
) for 5s
<5%U
T
(> 95% dip in
U
T
) for 0.5 cycle
40% U
T
(60% dip in U
T
)
for 5 cycles
70% U
T
(30% dip in U
T
)
for 25 cycles
<5%U
T
(>95% dip in
U
T
)for5s
Mains power quality
should by that of a typical
commercial or hospital
environment. If the user
of the Coronis 3MP
LED Display requires
continued operation
during power mains
interruptions, it is
recommended that
the Coronis 3MP LED
Display be powered from
an uninterruptible power
supply or a battery.
Power frequency (50/60
Hz) magnetic eld
IEC 61000-4-8
3A/m
Not applicable
Power frequency
magnetic elds should
be at levels characteristic
of a typical location in
a typical commercial or
hospital environment.
Conducted RF
IEC 61000-4-6
Radiated RF
IEC 61000-4-3
3Vrms
150 kHz to 80 MHz
3V/m
80 MHz to 2.5 GHz
3V
3V/m
Portable and mobile
RF communications
equipment should be
used no closer to any part
of the Coronis 3MP LED
Display, including cables,
than the recommended
separation distance
calculated from the
1. is the a.c. mains voltage prior to application of the test level.