34 Coronis Fusion 10MP
enabled. Please refer to “Ambient Light Compensation (ALC)” on
page 32 to enable ALC.
The available reading rooms for your C
ORONIS FUSION 10MP display are:
TO SELECT A READING ROOM:
1 Bring up the OSD main menu.
2 Navigate to the Configuration > Calibration > Ambient Light
menu.
3 Enter the Reading Room submenu.
4 Select one of the available reading rooms and confirm.
CR/DR/MAMMO Corresponds to light conditions in
diagnostic reading rooms for computed
radiology, digital radiology or
mammography. This setting has the lowest
maximum ambient light.
CT/MR/NM Corresponds to light conditions in
diagnostic reading rooms for computed
tomography, magnetic resonance or
nuclear medicine scans.
Staff Office Corresponds to light conditions in office
rooms.
Clinical Viewing
Room
Corresponds to light conditions in
diagnostic reading rooms for clinical
viewing.
Emergency Room Corresponds to light conditions in
emergency rooms.
Operating Room Corresponds to light conditions in operating
rooms. This setting has the highest
maximum ambient light.